X2
Synergy
MA X2.A Moisture Analyzer
MA X2.IC.A Moisture Analyzer
IMMU-21-12-02-19-EN
www.radwag.com
USER MANUAL

Table of Contents

Questions and Answers

  • K
    Kevin GrahamAug 6, 2025
    Why are my RADWAG Measuring Instruments measurements not repeatable?
    • A
      April TranAug 6, 2025
      Lack of measurement repeatability in RADWAG Measuring Instruments can stem from several factors: * Non-uniform sample content: Use a larger amount of substance to prepare the sample. * Insufficient drying time: Change the finish mode. * Excessive drying temperature causing sample oxidation: Lower the drying temperature. * Sample boiling during testing: Reduce the drying temperature. * Contaminated or defective temperature sensor: Clean the temperature sensor. * Unstable weighing table: Relocate the device to a more stable workstation. * Incompatible ambient conditions (vibrations, drafts): Adjust the ambient conditions to meet the guidelines specified in this service manual.

RADWAG MA X2.IC.A Series Specifications

General IconGeneral
CategoryMeasuring Instruments
ManufacturerRADWAG
TypeAnalytical Balance
Readability [d]0.1 mg
Linearity±0.2 mg
AdjustmentInternal
DisplayLCD
Power supply100 - 240 V AC 50/60 Hz
ConnectivityRS-232, USB, Wi-Fi