9.7. Custom Units
Parameter allowing you to declare two custom units. Displayed custom unit value is a result
of calculation, where obtained in the course of measurement weight value is multiplied by a
multiplier determined for this particular custom unit. The custom units can be freely named
with use of 3 characters maximum. By default the custom units are marked as: [u1] –
custom unit 1 and [u2] – custom unit 2.
9.8. Zeroing
To zero mass indication press key. Zero indication and the following pictograms
are displayed: and .
Zeroing operation means determining new zero point, recognized by the weighing device
as precise zero. The instrument can be zeroed only when the indication is stable.
Caution:
Indication can be zeroed only within ±2% range of maximum capacity. If the zeroed value
is above ±2% of the maximum capacity, then the software displays a respective error
message.
9.9. Taring
To determine net weight value, load the weighing pan with a packaging, wait for a stable
indication and press key. Zero indication and the following pictograms are
displayed: Net and . Upon unloading of the weighing pan, the sum of tared masses
with minus sign is displayed. You can assign tare value to a particular product stored in
the database. In such case the assigned tare value is automatically acquired upon
selection of the given product.
Caution:
It is impossible to tare negative values. When you try to tare negative values, the
weighing instrument responds with an error message. In such a case, zero the indication
and repeat taring procedure.