The process is carried out in a way presented below. After 8 minutes a numeric keyboard
is displayed. Enter the moisture analyzer temperature that is displayed on the control
thermometer.
Press button to confirm, the second step of adjustment starts. The moisture analyzer
halogen lamp starts to operate and the drying chamber is heated as long as necessary to
provide specified temperature, which is next maintained for 8 minutes. After 8 minutes a
numeric keyboard is displayed. Enter the moisture analyzer temperature that is displayed
on the control thermometer.
Press button to confirm, the third step of adjustment starts. The moisture analyzer
halogen lamp starts to operate and the drying chamber is heated as long as necessary to
provide specified temperature, which is next maintained for 8 minutes. After 8 minutes a
numeric keyboard is displayed. Enter the moisture analyzer temperature that is displayed
on the control thermometer.
Press button to confirm.
Temperature adjustment process is completed and moisture analyzer displays
<Adjustment> window.
For moisture analyzers with maximum heating temperature of 250°C, the temperature
adjustment process is likewise, only the temperature values are higher for each part of
the process.
11.9. Temperature Test
The parameter allows to check the correctness of indications of moisture analyzer
temperature.