• If you fail to establish the communication, contact RADWAG service team.
The selected network and settings of parameters for connection are stored in moisture
analyzer memory. The moisture analyzer connects to the network in accordance with
these settings each time it is switched on.
To disconnect the network turn the communication off:
COMMUNICATION/WIFI/WIFI – NO
21.4. USB Ports
USB 1 port of type A is intended for:
− connecting a USB flash drive storing <FAT files system>,
− connecting moisture analyzer to PCL printer,
− connecting EPSON TM-T20 printer to USB port.
USB flash drive may be used for printing data on measurements (setting PRINTER/PORT
parameter to PENDRIVE value). Measurement data printout operation is described in
section 21.3. of this service manual. Additionally, databases (for description of this function
go to section 14 of this service manual) and user’s parameters settings can be copied
balance to balance. Upon inserting USB flash drive, an Export/Import group of parameters
opens automatically. If you are authorized you can:
Export: databases, user’s parameters
Import: databases, user’s parameters
During export the software saves files with databases and parameters content to a flash
drive. The exported data can be imported to any X2 series balance.
Please remember that in case of PCL printer, completely filled page is printed. This means
that in order to print the page, it is necessary to press PRINT key, located on the operation
panel, for several number of times (printout size determines how many times the PRINT key
must be pressed prior printout).
It is possible to obtain printout upon pressing PRINT key once only, supposing that control
code <0C> has been set as SUFFIX (for description of this function go to section 21.2
'Printer').
USB 2 port of type B is intended for:
− connecting moisture analyzer to a PC.
In order to make connection of moisture analyzer and computer possible, it is necessary to
install virtual COM port in a computer.
To carry out this procedure, you need a respective driver installer which may be either
downloaded from the website www.radwag.pl or taken from a CD storing the manuals: R X2
SERIES RADWAG USB DRIVER x.x.x.exe.
Steps:
1. Run driver installer.