24.3. Manual Printout / Automatic Printout
The moisture analyzer enables generating manual or automatic printouts.
• Manual printout: upon indication stabilization press key.
• Automatic printout is generated automatically in accordance with the settings for
automatic printout (read section 9).
The content of printout depends on settings of <Standard printout> menu - <GLP
Printout> (read section 12.3).
Mass printout format:
space stability mass space unit CR LF
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
[space] for positive values
[-] for negative values
9 characters with decimal point, right justification
3 characters, left justification
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - a printout generated, with reference to <GLP
printout> settings, from a weighing device upon pressing key: