Integration with PUE 7.1, PUE 10 indicator
Format: NT CR LF
command not recognised (mass format incorrect)
MASS FRAME
- command carried out, immediate response: mass value in basic measuring
unit
1
2
3
4
5
6
7
8
9-18
19
20-22
23
24-32
33
34-36
37
38
39
40
N
T
space
Stability
marker
space
Mass
space
Mass unit
space
tare
space
Tare unit
space
[space] if stable, [?] if unstable
[space] for any value but zero, [Z] for zero value
marker informing on range within which mass value is comprised:
[space] I range balance, [2] II balance, [3] III balance
[zero] no digit, [1] one digit marker, [2] two digits marker, [3] three digits
marker, [4] four digits marker, [5] five digits marker
10 characters for net mass given in adjustment unit (with dot and a
floating-point ā-ā sign ) with right justification
3 characters, left justification
9 characters with dot - right justification (when floating-point tare gets
turned off automatically then zero value is sent)
3 characters, left justification
number specifying quantity of hidden digits: [space] when no digits are
hidden, [1] when 1 digit is hidden
command sent from a computer
N T _ ? _ _ 0 _ _ _ _ _ - 5 . 1 1 3 _ g _ _ _ _ _ _ _ 0 . 0 0 0 _ g _ _ _0 CR LF
- command carried out, response: mass value and remaining data