Moisture analyzer display shows name and software number first, next an indication 0.000 g
(moisture analyzers with readability of 1 mg) or 0.0000 g (moisture analyzers with readability
of 0.1 mg). If the indication is different than zero, press key.
4.7. Temperature Stabilization Time
Before start of the measuring processes, it is necessary to wait until the moisture analyzer
reaches thermal stabilisation.
Moisture analyzer stored in much lower temperatures, than the workroom temperature,
before being plugged to the mains (e.g. during winter period) must be subjected to thermal
stabilisation. The thermal stabilisation period takes about 4 hours. During the thermal
stabilization, the indications on the screen can change. Potential workroom temperature
change shall occur gradually and slowly in the course of the weighing instrument operation.
4.8. Connecting Peripheral Equipment
Use only accessories and peripheral equipment recommended by the manufacturer. The
moisture analyzer must be disconnected from the mains before connecting or disconnecting
any peripherals (printer, PC computer, computer keyboard of USB type, additional display).
Upon connecting the peripherals, plug the moisture analyzer to the mains.