- 42 -
Upon indication stabilization, packaging mass is automatically tared (Net
pictogram is displayed in the upper part of the display).
Put/pour/dispense product that is to be packed into the packaging.
Net mass is displayed.
Unload the weighing pan (remove both product and packaging).
When the weighing device detects weight value lower than gross weight value set
in <AUTO THRESHOLD> parameter, the entered tare value gets automatically
deleted (Net pictogram is no longer displayed).
Load the weighing pan with packaging of another product, upon indication
stabilization, packaging mass is automatically tared (Net pictogram is displayed in
the upper part of the display).
Put/pour/dispense product that is to be packed into the packaging.
For correct operation of the weighing instrument with AUTOTARE function, it is necessary to
adjust the threshold value.
<AUTO THRESHOLD> parameter refers to the following functions:
- autotare,
- automatic printout.
No automatic taring takes place as long as the gross weight value stays within the
range set in <AUTO THRESHOLD> parameter.
9.14. Print Mode
Function designed to enable print mode setting, i.e. key activation.
Available values:
<WHEN STABLE> - for this value stable measurement result, along with the settings
for <GLP PRINTOUT> parameter, is sent to the printer port. Upon pressing key
at the moment when the result is unstable (no pictogram displayed), the
program first waits for the stability condition to be met, next sends the
measurement result to the port.
<EACH> - every single pressing of key results with sending the measurement
indication to the printer port along with the settings for <GLP PRINTOUT>
parameter. Every single indication is sent (stable and unstable). In case of unstable
indication, <?> sign is displayed in front of the weighing result frame.
Function available for non-verified moisture analyzers exclusively.
<AUTO> - select this value to enable automatic printout. When selecting <AUTO>
value remember to set <AUTOPRINT THRESHOLD> in accordance with your
requirements.
<AUTO + INTERVAL> - select this value to start automatic print and record of
indications in WEIGHINGS and ALIBI databases, carried out in a cyclic manner in a
specified time intervals. The interval is set in [min] in <INTERVAL> parameter.
Interval range is 1 - 9999 min.
Caution:
Each weighing result is printed and recorded (stable and unstable for a non-verified
moisture analyzer, stable for a verified moisture analyzer).
Automatic operation with interval starts at the moment of switching the function on.
The first stable weighing result of value greater than AUTO THRESHOLD value is
printed and recorded as the first measurement. The following measurements are
printed with frequency set in INTERVAL parameter. To stop the automatic operation
with interval, switch the option off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG MA X2.IC.A Series and is the answer not in the manual?

RADWAG MA X2.IC.A Series Specifications

General IconGeneral
BrandRADWAG
ModelMA X2.IC.A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals