- 43 -
Upon switching the function of auto print with interval on, PRINT button becomes
inoperative (no indication is printed when pressed).
Automatic operation procedure:
Press key to zero the indication (stability marker , and zero pictogram
are displayed).
Load the weighing pan, the first stable measurement is sent to the printer port.
Unload the weighing pan.
The next measurement is possible when the indication gets lower than the value
set for <AUTO THRESHOLD> parameter (next measurement does not require
zero value).
For automatic operation it is necessary to adjust the threshold value. The
measurement is sent to a computer or a printer when mass indication gets below the
set net value of <AUTO THRESHOLD> parameter.
<AUTO THRESHOLD> parameter refers to the following functions: automatic tare,
automatic operation, autoprint with interval.
9.15. Minimum Sample Weight
Weighing mode comprises <Minimum sample weight> function. In order to use this function
it is necessary to enter minimum sample weight (MSW) value and tare values for which the
MSW value is to be obligatory. For standard MAX2 series the values are zero.
Only authorized RADWAG employee or operator with Administrator permissions,
providing that moisture analyzer factory settings enable this, can carry out procedure
aiming to determine minimum sample weight and next enter the respective data.
Should you use the minimum sample weight function, and need data on minimum sample
weight to be entered to moisture analyzer menu, please contact the local representative of
“RADWAG” company.
Using reference sample weights, the authorized RADWAG employee determines minimum
sample weight for specified tare containers. This operation is carried out at the place of use
and in accordance with requirements of the applied quality system. The obtained value is
entered into the weighing instrument program.
The weighing instrument program enables defining tare value with assigned minimum
sample weight value.
<Minimum sample weight> function guarantees that the results of weighing operation are
comprised within the set tolerance, in accordance with applied quality management system
of particular company.
Caution: Function valid for weighing mode exclusively.
Available options:
MODE
Noneminimum sample weight function off.
Block select to enable display of respective pictograms informing you about the
measured mass (whether it is lower than the MSW value or higher than the MSW
value); with this option on, the program makes it impossible to accept the measurement
that is lower than the MSW value.
Warnselect to enable display of respective pictograms informing you about the
measured mass (whether it is lower than the MSW value or higher than the MSW
value). You can accept the measurement that is lower than the MSW value.
Taremaximum tare value, for which the minimum sample weight is obligatory
(analyse the below examples).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG MA X2.IC.A Series and is the answer not in the manual?

RADWAG MA X2.IC.A Series Specifications

General IconGeneral
BrandRADWAG
ModelMA X2.IC.A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals