<Automatic adjustment time> parameter determines time interval between successive
automatic internal adjustments. The time interval is declared in hours and ranges between
0.5 and 12 hours.
To set time interval for automatic adjustment:
select <Automatic adjustment time> option,
using displayed menu, select appropriate time interval (given in hours) which is to
elapse between the most recent adjustment and activation of the following automatic
internal adjustment.
11.7. Report Printout
Adjustment report is automatically generated at the end of each adjustment process or
adjustment test, next it is sent to port selected for <PERIPHERALS/PRINTER> (COM 1
port by default). To declare report content go to <PRINTOUTS/ADJUSTMENT REPORT>
submenu.
For instruction on how to declare adjustment report settings read section 'Printouts'.
The report can be printed using printer connected to the moisture analyzer or sent to a
computer and saved as a file for the archiving purposes.
11.8. Adjustment of the Drying Chamber Temperature
Additional equipment is required for temperature adjustment.
1. Thermometer.
2. Thermometer holder with shield.
Adjustment process is carried out to adjust the temperature of drying chamber sensor. In
order to carry out adjustment of moisture analyzer temperature, place the control
thermometer as it is presented below: