2.10. Protective Clothing
While working with the instrument use protective clothing, this is to take safety precautions
against potential hazards source of which might be tested samples and ingredients.
Use the following while carrying out tests:
protective apron,
protective glasses,
protective gloves (while working with hazardous chemical substances).
Before the use of the above-mentioned protective clothing, make sure that it has been
designed to be used with specific samples and that it is not damaged.
3. TRANSPORT AND STORAGE
3.1. Delivery Check
Upon delivery it is necessary to check the package and the device, make sure that your
package bears no signs of damage.
3.2. Packaging
Keep all package elements should your device be transported in the future. Remember
that only original packaging can be used for shipping purposes. Prior packing, uncouple
any cables, remove any separable components (weighing pan, shields, inserts). The
device components must be packed into an original packaging providing protection
against potential damage during transportation.
4. UNPACKING AND INSTALLATION
4.1. Place of Use
Operate the device in a workroom free of vibrations and shakes, where there are no
air drafts nor dust. The workstation has to be located up to 2000 m above sea level.
When setting up the moisture analyzer leave enough space to prevent heat from
building up and to keep your analyzer from overheating. Leave about 20 cm around
the instrument and about 1 m above.
Make sure that the ambient temperature ranges between: +10
°
C ÷ +40
°
C.
Relative humidity cannot exceed 80% at the temperature of 31
°
C, decreasing
linearly to 50% of the relative humidity at the temperature of 40
°
C.
Place the moisture analyzer either on a robust-design table or on a wall bracket,
which is both distant from heat sources and insusceptible to vibrations.
Take special precaution while weighing magnetic objects, as part of the moisture
analyzer is a strong magnet.