Caution:
For Step profile set step temperature „1” and „2”, drying temperature and heating time for
step 1 and 2. Heating time is measured from the moment of reaching the set temperature for
respective step.
18.2. Finish Mode
Finish mode parameter's values:
Automatic 1 - auto switch-off (1mg /10s),
Automatic 2 - auto switch-off (1mg /25s),
Automatic 3 - auto switch-off (1mg /60s),
Automatic 4 - auto switch-off (1mg /90s),
Automatic 5 - auto switch-off (1mg /120s),
Manual- manual switch-off (upon pressing Start/Stop, maximum drying time is 99 hours
59 minutes). Upon the elapse of pre-set time the moisture analyzer finishes drying process.
Time-defined- time-defined switch-off (maximum drying time is 99 hours 59 minutes).
Defined 1 - user-defined switch-off (change of mass comprised within the moisture
analyzer range, set with moisture analyzer readability, time interval ranging from 1 to
120s).
Defined 2 - user-defined switch-off (change in humidity set with moisture analyzer
readability, time interval ranging from 1 to 60s).
Test - (enables selection of the auto switch-off parameters for a given sample).
When drying using auto switch-off option, the printout contains information about time and
results of respective auto switch-off cases. You can use them to determine required switch-
off option (result and time).