EasyManua.ls Logo

RADWAG MA X2.IC.A Series

RADWAG MA X2.IC.A Series
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 77 -
Caution:
For Step profile set step temperature „1and „2”, drying temperature and heating time for
step 1 and 2. Heating time is measured from the moment of reaching the set temperature for
respective step.
18.2. Finish Mode
Finish mode parameter's values:
Automatic 1 - auto switch-off (1mg /10s),
Automatic 2 - auto switch-off (1mg /25s),
Automatic 3 - auto switch-off (1mg /60s),
Automatic 4 - auto switch-off (1mg /90s),
Automatic 5 - auto switch-off (1mg /120s),
Manual- manual switch-off (upon pressing Start/Stop, maximum drying time is 99 hours
59 minutes). Upon the elapse of pre-set time the moisture analyzer finishes drying process.
Time-defined- time-defined switch-off (maximum drying time is 99 hours 59 minutes).
Defined 1 - user-defined switch-off (change of mass comprised within the moisture
analyzer range, set with moisture analyzer readability, time interval ranging from 1 to
120s).
Defined 2 - user-defined switch-off (change in humidity set with moisture analyzer
readability, time interval ranging from 1 to 60s).
Test - (enables selection of the auto switch-off parameters for a given sample).
When drying using auto switch-off option, the printout contains information about time and
results of respective auto switch-off cases. You can use them to determine required switch-
off option (result and time).

Table of Contents

Related product manuals