EasyManua.ls Logo

Rae QRAE 3 - Calibration Procedures; Zero Calibration; Span Calibration for Pumped Models

Rae QRAE 3
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
10.3. Zero Calibration
This operation sets the zero point of the sensor calibration curve for clean air. It should be
performed before other calibrations.
Note: If you use a zero air cylinder, you must use the QRAE 3 Calibration Adapter (and a T
calibration tube if the instrument has a pump). Using a calibration adapter is not necessary for
calibration in fresh air.
10.3.1. Zero Calibration
This procedure determines zero points of most sensors. The QRAE 3 should be zero-calibrated in
clean air with 20.9% oxygen or with a cylinder of clean zero air.
At the Calibration menu, select “Fresh Air” by pressing [Y/+] once to enter fresh air calibration.
After a timer countdown, the zero calibration is done. The LCD displays the sensor names and
tells you whether each calibration passed or failed, followed by the sensor readings.
Note: You can abort the calibration at any time during the countdown by pressing [MODE].
10.3.2. Single-Sensor Zero Calibration
Select the sensor and then start the calibration by pressing [Y/+]. You can abort the procedure
anytime by pressing [MODE].
10.4. Span Calibration
This procedure determines the second point of the sensor calibration curve for the sensor.
Note: When a manual calibration is performed, the readings shown are in the equivalent
units of the calibration gas, and not the measurement gas.
10.4.1. QRAE 3 Pumped Model
The QRAE 3 pump draws at a flow rate of between 200cc/min and 450cc/min. The
instrument must be connected to a cylinder of calibration gas with supplied tubing
featuring a T calibration tube (P/N M02-3008-000), as illustrated below.
Note: A constant-flow regulator with flow rates from 500cc/min to 1000cc/min should be used.
Note: Make sure pressure in
the calibration gas cylinder
is higher than 100 psi when
using a T calibration tube.

Other manuals for Rae QRAE 3

Related product manuals