Table A: Sensor Conguration
Sensors are not interchangeable. Use only RAE Systems
sensors, and use only the sensor type specied for your ToxiRAE
II monitor. Use only RAE Systems batteries. Use of non-RAE
Systems components will void the warranty and can compromise
the safe performance of this product.
Model and
Part Number
Range
(ppm)
Resolution
(ppm)
Span
Gas
(ppm)
Low
(ppm)
High
(ppm)
STEL
(ppm)
TWA
(ppm)
Sensor
Warm-Up
Time **
NH
3
045-0518-000
0-50 1 50 25 50 35 25 20 min
CO
045-0512-000
0-500 1 100 35 200 100 35 20 min
CO*
045-0512-200
0-1999 10 100 35 200 100 35 20 min
Cl
2
*
045-0516-000
0-10
0.1 10 0.5 5 1 0.5 20 min
ClO
2
*
045-0523-000
0-1
0.01 0.5 0.2 0.5 0.3 0.1 20 min
HCN*
045-0517-000
0-100 1 10
4.7 50 4.7 4.7 20 min
H
2
S
045-0511-000
0-100 1 25 10 20 15 10 20 min
NO*
045-0514-000
0-250 1 25 25 50 25 25 4 hours
NO
2
*
045-0515-000
0-20
0.1 5 1 10 1 1 20 min
O
2
*
045-0006-000
0-30
0.1
0.0%
(N
2
)
19.5 23.5 - - 20 min
PH
3
*
045-0519-000
0-5
0.01 5 1 2 1 0.3 20 min
SO
2
*
045-0513-000
0-20
0.1 5 2 10 5 2 20 min
*Call for availability
** Applies to newly installed sensor only