MTR221/A - MTR221/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R221/A - R221/A DHHS
SPECIFICATIONS
9
R221/A - R221/A DHHS - Specifications
Ralco srl
SID: (optional)
Precision of Measurement with Retractable Tape
EN 60601-2-54
< 2% SID
Maximum Radiation Leakage
Measured at 100 cm with X-ray Beam = 150 kVp - 4 mA
EN 60601-2-54
< 40 mRh
Potentiometers
Square Field
Round Field
n.a.
n.a.
Maximum Load for Accessory Guides
GUIDES FOR ACCESSORIES (IF ASSEMBLED): PRIOR TO INSERTING
ACCESSORIES IN THE GUIDES CHECK ON THE PERFECT ATTACHMENT
OF THE SUPPORT WITH THE MOUNTING SLOT ON THE COLLIMATOR
(
TOLERANCE MAX. ± 0,5 MM). A FAULTY ATTACHMENT COULD BE
DANGEROUS AND IT COULD CAUSE THE FALL OF ACCESSORIES
Static load: 70 N (about 7.1 Kg)
Dynamic load: 15 Nm (approx.
3.06 Kg)
Operation Environment
• Ambient Temperature
• Relative Humidity
• Atmospheric Pressure
10°- 40°C
0%- 75%
700 - 1060 hPa
Storage
• Temperature
•Humidity
• Atmospheric Pressure
-40° - 70°C
10% - 95%
500 - 1060 hPa
Weight 7.7 Kg
Dimensions
L: 270.5 mm
W: 183 mm
H: 168 mm