After all the necessary snapshots are taken, calibration module will start automatically.
Resulting accuracy of calibration is specified in pixels. A good result is accuracy no
worse (not higher) than 0.1 pix. If you get a higher value, repeat the calibration, carefully
making sure that the plate is sharp, bright and correctly positioned.
Someadviceoncalibration:
● donotchangethedistancefromthecenteroftheplatetothescannerwhen
turningtheplate.Theexceptionareonlysnapshotsinpositions10and11,
● donotrotatetheplateataveryhighangle.Ensurethatallmarksontheplateare
visiblefrombothcameras,whenyouturnthetable,
● becarefulwithcalibrationplates!Itisnotpermittedtocontaminateorperform
anymechanicaldamagetothesurfacewiththemarks.Aftertheusestorethe
platesindesignatedcase,
● beforethecalibrationprocedureensurethatcamerasandcablesaresecurein
thedesignatedposition.
Fastcalibration(orientation)