User Guide of RFL-C3300W~RFL-C15000W
38
Figure 22 RAYCUS Client Software Alarm Log interface
5.2 Alarm Solutions
Table 6 Alarms and possible solutions
Alarm message
Error instructions and possible solutions
Temperature
Error
Instruction:
Low temperature/high temperature alarm, occurs when the
temperature at the monitoring point exceeds the pre-set
upper/ lower limit.
Solutions:
High temperature alarm:
Check if the water-cooling system is working normally, if
the water temperature is set correctly, or if the water
pipe-line connection is correct.
When the water cooling system works normally and the
water temperature drops to below 30°C, restart the laser to
see if the problem is solved.
If the alarm continues to happen, please contact Raycus.
Low temperature alarm:
Check if the water temperature is too low or not.Besides, a
low ambient temperature may also cause a low temperature
alarm when the laser is cold started. Please wait till the water
temperature rises to above 10°C and then restart the divice.
If the alarm continues to happen, please contact Raycus.