EasyManua.ls Logo

Re T-one ULS - Quick Start-up Guide; Installation Procedure; Adjustment Mode Entry; Initial Calibration Steps (Part 1)

Re T-one ULS
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
T-one ULS
Rev. 06/18 10/26
Quick start-up
This section describes the installation and test procedures that must be carried out if you wish to set the
instrument up quickly.
A more detailed description of how each function works is given in § Detailed description of programming
functions (page 16).
Installation
1. Fit the T-one ULS regulator.
2. Make the electrical connections as illustrated in § Electrical connection diagrams (pages 12-13).
3. Connect the instrument to a power supply.
Adjustment mode
4. Press the enter key and keep pressed for about one second, to enter the programming environment;
the display shows F.1.
5. Use the up key to select function 4.
6. Press the enter key: the word “CAL” flashes on the display for a few seconds; then, the calibration
functions needed for the quick start-up of the instrument are given in sequence.
Note: During calibration the instrument can not perform the states switch, neither by frontal nor by
remote, except for PRIORITY STOP.
Calibration functions
7. F.30 appears on the display: use this function to set the analog input from which the “reference
diameter” will be acquired
press enter and the function value set is displayed;
using the up or down keys set the value to 0 to use the AIN1 input (0÷10Vdc) or 1 to use the
AIN0 input (4÷20mA);
press enter to confirm and up to move onto the next function.
8. F.31 appears on the display: use this function to set the position of the decimal point in the diameter
displaying;
press enter and the function value set is displayed;
using the up or down keys set the value to 0 to have not the decimal point (0000), 1 to have one
digit after the decimal point (000.0) or 2 to have two digits after the decimal points (00.00);
press enter to confirm and up to move onto the next function.
9. F.32 appears on the display: use this function to set the maximum diameter value;
press enter and the function value set is displayed;
using the up or down keys set the maximum diameter value;
press enter to confirm and up to move onto the next function.
10. F.33 appears on the display:
position the reel with the maximum diameter on the unwinder/rewinder, position the diameter
sensor correctly and calibrate it, if necessary (if using the US3 sensor, see the relative user
manual);
press enter, the word “YES” flashes on the display for a few seconds if the gain calibration was
successful;

Table of Contents

Related product manuals