Released 2018-01-19 Drawing No. LP0722
USER INPUT ASSIGNMENT
Select the value(s) to which the User Input Function is assigned. The
User Input Assignment only applies if a selection of reset, or display hold
is selected in the User Input Function menu.
MODEDISPLAY
No Function
NO
DESCRIPTION
User Input disabled.
Program Mode Lock-out
P-Loc
Zero Input
(Edge triggered)
ZErO
Zero the Input Display value causing
Display Reading to be Offset.
d-SEL
d-LEV
d-HLd
rESEt
Resets the assigned value(s) to the current
input value.
HI-LO HI
dSPLO
See Programming Mode Access chart
(Module 3).
Setpoint 1 and 2 Reset
Setpoint 1 Reset
Setpoint 2 Reset
Reset both setpoint 1 and 2 outputs.
Resets setpoint 1 output.
Resets setpoint 2 output.
rSt12
rSt-1
rSt-2
USER INPUT FUNCTION
NO
USrIN
dSP
U-ASN
Display Select
(Edge Triggered)
Display Intensity Level
(Edge Triggered)
Display Hold
Reset (Edge triggered)
Advance once for each activation.
Increase intensity one level for each
activation.
Holds the assigned display, but all other
meter functions continue as long as
activated (maintained action).
LO
U-Act
USER INPUT ACTIVE LEVEL
Select whether the user input is configured as active low or active
high.
LO HI
Entering Code 66 will overwrite all user settings
with the factory settings. The meter will display rESEt
and then return to CodE 00. Press the PAR button to
exit the module.
MIN DISPLAY ENABLE
YESNO
NO
LO-En
NO
FCS
2.0
LO-t
MIN CAPTURE DELAY TIME
When the Input Display is below the present MIN value for the entered
delay time, the meter will capture that display value as the new MIN
reading. A delay time helps to avoid false captures of sudden short
spikes.
0.0 to 999.9 sec.
Select yES to perform any of the Factory Service Operations shown
below.
FACTORY SERVICE OPERATIONS
yESNO
Enables the Minimum Display Capture capability.
The PAXLA uses stored calibration values to provide
accurate measurements. Over time, the electrical
characteristics of the components inside the PAXLA
will slowly change, resulting in the stored calibration
values no longer accurately defining the input circuit.
For most applications, recalibration every 1 to 2 years should be sufficient.
Calibration of the PAXLA involves a calibration which should only be
performed by individuals experienced in calibrating electronic equipment.
Allow 30 minute warm up before performing any calibration related
procedure. The following procedures should be performed at an ambient
temperature of 15 to 35 °C (59 to 95 °F).
CAUTION: The accuracy of the calibration equipment will directly
affect the accuracy of the PAXLA.
CALIBRATION
66
CodE
RESTORE FACTORY DEFAULT SETTINGS
Entering Code 50 will display the version (x.x) of
the meter. The display then returns to CodE 00.
Press the PAR button to exit the module.
50
CodE
VIEW MODEL AND VERSION DISPLAY
48
CodE
5.2 Module 2 - secondary FunctIon paraMeters (2-SEC)
PAR
Max Capture
Delay Time
Max Display
Enable
Min Display
Enable
Access Code
For Service
Operations
Min Capture
Delay TIme
Factory
Service
Operations
2-SEC
HI-En HI-t
LO-En LO-t FCS
CodE
Pro
PARAMETER MENU
2.0
HI-t
MAX CAPTURE DELAY TIME
When the Input Display is above the present MAX value for the
entered delay time, the meter will capture that display value as the new
MAX reading. A delay time helps to avoid false captures of sudden short
spikes.
0.0 to 999.9 sec.
NO
HI-En
MAX DISPLAY ENABLE
YESNO
Enables the Maximum Display Capture capability.