52
c) Press the release button
on the lancing device to
lance the skin.
Continue to
hold the lancing device
firmly against the skin until
a blood drop forms
.
d) Once a
large enough
drop of blood
has formed,
remove the
lancing device.
Step 4:
Pick up the meter and touch the end of the test
strip to the blood sample until the reaction site is
full. Immediately remove the meter and test strip
from the blood drop.
IMPORTANT
Repeat blood draw if fluid is clear.
IMPORTANT
If it takes longer than 20 SECONDS to obtain
a blood sample, and any sample has been
applied to the strip, discard strip and use a
new strip before proceeding.