2.3
RMI-Q basics
P1, P2, P3, P4 SYSTEM STATUS LEDs
Off - Probe number empty.
Yellow - Probe number full, in standby.
Green - Probe number full, operating.
Red - Probe number full, 0.5 s
compatibility error.
Yellow/off - Flashing: Selection error.
Green/yellow - Flashing: Acquisition/clearing
pending, full.
Green/off - Flashing: Acquisition pending,
empty.
Red/yellow - Flashing: Acquisition complete,
probe full.
Red/off - Flashing: Clearing complete,
probe empty.
LOW BATTERY / START LED
Off - Battery is OK and no M code
start/stop in progress.
Red - Battery is low.
Green - M code start/stop in progress.
Yellow - Battery low and M code
start/stop in progress.
Red/off - Flashing: RMI-Q has an
overcurrent condition.
PROBE STATUS LED
Red - Probe triggered.
Green - Probe seated.
Red/off - Flashing: RMI-Q has an
overcurrent condition.
ERROR LED
Off - No error.
Red - Error, other outputs may be
incorrect.
Red/off - Flashing: RMI-Q has an
overcurrent condition.
SIGNAL LED
Off - No probes operating.
Green - Good communications.
Green - yellow - Good communications.
Red - Poor communications, radio link
may fail.
Red/off - Flashing: RMI-Q has an
overcurrent condition.
NOTES:
The 'PROBE STATUS' LED is always illuminated
when power is present at the RMI-Q (as the
RMI-Q does not incorporate a separate ‘power
present’ indicator).
All LEDs report the status of the partner RMP. If
there is no partner in range, or the partner is off,
then the 'PROBE STATUS' and 'ERROR' LEDs
will illuminate red. The 'LOW BATTERY/START'
and 'SIGNAL' LEDs will be off.
When the RMI-Q is powered on it will enter
the acquire partner mode. This is indicated by
a flashing green 'SIGNAL' LED (no change in
outputs). After ~60 seconds it will switch to its
normal mode listening for its partner.
The conditions shown by the 'LOW BATTERY/
START', 'PROBE STATUS' and 'ERROR' LEDs
are the same as those present on the electrical
signal outputs.