All Primary Screen Buttons are displayed to navigate to those screens.
Alarms, Trending, Batch Info, Set Up, Administration
When Starting the Diaphragm Pump, the following dialog box will appear:
Yes setting is preferred when stopping the Diaphragm Pump and restarting with the same flow
rate or process settings (i.e. same PRV and PV values as when the Diaphragm Pump last ran).
Default setting is preferred when starting the Diaphragm Pump with new flow rate settings or
new process setup (e.g. different bioreactor configuration & parameters). This minimizes the
number of cycles taken by the controller to reach the desired flow rate. At any selected flow rate,
default simply resets the control parameters to factory preset values.
Cancel will return to the main screen without any action taken.
When Stopping the Diaphragm Pump, the following dialog box will appear: