English 1
Disinfection and sterilisation guide Clinical use only
This guide is intended for multipatient use of the AirFit
™
F20/ AirFit
™
F20 for Her ResMed full face mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for
cleaning instructions. This guide describes ResMed’s recommended and validated procedures for cleaning, disinfection and sterilisation of the mask in accordance with ISO17664.
AirFit F20 / AirFit F20 for Her
mask component
1
Thermal disinfection Chemical disinfection Sterilisation Validated number of
cycles
2
Manual AWD
3
CIDEX
™
OPA Gigasept FF STERRAD
™
100S STERRAD
™
NX
• Cushion 30
• Headgear
– – –
30
• Elbow
– – –
30
• Frame
– – – –
30
1
This mask may not be available in all regions. For full details regarding the correct use of these masks, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Components Card on www.resmed.com.
2
If a healthcare facility requires an additional disinfection or sterilisation cycle after reassembly, the number of validated cycles must be halved.
3
Automatic Washer-Disinfector
A Elbow
1 Valve (AAV)
2 Vent
3 Side buttons
4 Swivel
B Frame
C Cushion
D Headgear
5 Magnetic clips
6 Lower headgear straps
7 Upper headgear straps
A+B+C Frame system
A+B+C+D Complete system