AirSense 10 and AirCurve 10 reprocessing guide
This guide is intended for multipatient use of the AirSense™10 and AirCurve™10 devices.
When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health care provider, the
cleanable water tub, air outlet and air tubing should be reprocessed between each patient use.
If the cleanable water tub or the air tubing are being used for a single user in the home, refer to the cleaning instructions in this
guide or in the User Guide.
Described here are ResMed’s recommended and validated procedures for cleaning and disinfecting the cleanable water tub, air
outlet and air tubing. However, the steps for disinfection vary regionally and each healthcare facility should consult its own
procedures before carrying out those within this guide.
Note: The standard water tub cannot be disinfected. If contaminated, it must be discarded and replaced with a new water tub.
WARNING
ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their effect on the
performance of the product, will be acceptable.
When using detergents, disinfectants or sterilization agents, always follow the manufacturer's instructions.
Beware of electrocution. Do not immerse the device, power supply or power cord in water. If liquids are spilled into
or onto the device, unplug the device and let the parts dry. Always unplug the device before cleaning and make sure
that all parts are dry before plugging it back in.
Surface disinfection
1. Wipe the exterior of the device including display, externally accessible ports, side cover, power supply unit and accessories
with a disposable cloth and mild detergent or alcohol disinfectant (see list below).
2. Remove any excess disinfectant with a disposable dry cloth.
Agents recommended for surface disinfection and cleaning:
Warm water and mild detergent eg, Teepol
multipurpose detergent
Window cleaner or other premixed surface detergent
Methyl alcohol solution
70% Ethyl alcohol solution
70-90% Isopropanol solution
10% Bleach solution
Isopropyl wipes
CaviCide
Mikrozid
®
Actichlor
Plus
Terralin
®
.
Note: Agents may not be available in all regions.

Other manuals for ResMed AirSense 10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirSense 10 and is the answer not in the manual?

ResMed AirSense 10 Specifications

General IconGeneral
Pressure Range4-20 cm H2O
Ramp Time0 to 45 minutes (5-minute increments)
Therapy ModeCPAP, AutoSet, AutoSet for Her
Data StorageSD card
Altitude AdjustmentAutomatic
Leak CompensationYes
Device TypeContinuous Positive Airway Pressure (CPAP) Machine
Dimensions116 mm x 255 mm x 150 mm
Weight1248 g
HumidificationIntegrated humidifier
Power Supply100-240V, 50/60Hz
Heated TubingOptional ClimateLineAir
ConnectivityWireless
Power Supply Voltage100-240V
Connectivity OptionWireless
Climate ControlAutoSet Climate Control

Related product manuals