EasyManuals Logo

ResMed AutoSet CS 2 User Manual

ResMed AutoSet CS 2
79 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
20
An anti-bacterial filter with high impedance can cause safety cutouts in the
A
UTO
S
ET
CS2
if it is in use during the combination of:
high peak mask pressure (EEP + Max PS = 20 cmH
2
O)
high peak patient inspiration (>50 L/min)
significant mask leak flow (>50 L/min).
It is recommended that an anti-bacterial filter with a pressure drop less than or equal
to 1 cmH
2
O at 60 L/min flow is used if this combination of conditions is present.
SUPPLEMENTAL OXYGEN
The
AUTOSET CS2
is designed to be compatible with the use of low flow oxygen
therapy provided from an external source and supplied at a fixed rate of up to 15 L/min.
Actual oxygen concentrations will vary depending on the mask used, where the oxygen
is introduced, pressure setting, volume delivered, leak, and patient breathing pattern.
!
WARNING
If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating.
Explanation: When the device is not in operation, and the oxygen flow is left
on, oxygen delivered into the ventilator tubing may accumulate within the device
enclosure and create a risk of fire.
Always begin AUTOSET CS2 therapy before the oxygen supply is turned on.
Always turn the oxygen supply off before stopping AUTOSET CS2 therapy.
Oxygen supports combustion. Oxygen should not be used while smoking or in
the presence of an open flame.
26853r2clin.book Page 20 Wednesday, October 14, 2015 4:03 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AutoSet CS 2 and is the answer not in the manual?

ResMed AutoSet CS 2 Specifications

General IconGeneral
BrandResMed
ModelAutoSet CS 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals