Recommended separation distances between portable and mobile RF communications
equipment and the
AutoSet CS2
The AutoSet CS2 is intended for use in an environment in which radiated RF disturbances are controlled. The
customer or the user of the AutoSet CS2 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the AutoSet
CS2 as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output
power of transmitter
W
150kHz to 80MHz
d = 1.17 P
80 MHz to 800 MHz
d = 0.35 P
800MHz to 2.5 GHz
d = 0.70 P
0.01 0.17 0.04 0.07
0.1 0.370.110.22
1 1.170.350.70
10 3.69 1.11 2.21
100 11.70 3.50 7.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
26853r2eng.book Page 72 Wednesday, October 14, 2015 4:08 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AutoSet CS 2 and is the answer not in the manual?

ResMed AutoSet CS 2 Specifications

General IconGeneral
BrandResMed
ModelAutoSet CS 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals