Mute Light indicator
Configuration/Blinking pattern Mute Light indicator
Normal Operation (Unmuted)
Normal Operation (Muted)
Special features
Drop detection
In the event that the Multi Mic falls from a height of more than
75 cm (30 inches), a built-in sensor will mute the output streamed
to the hearing instruments for up to 2 seconds. This is done to
protect the hearing instrument user from very loud and unpleas-
ant impact noise should the unit be accidentally dropped during
use.
Volume control lock
If needed, the volume of the Multi Mic can be locked to a specific
level.
The volume level can be locked for Microphone and FM mode
only. When the volume control is locked, it will be fixed to the level
it had when the Multi Mic was powered OFF, but it is still possible
to use the mute/unmute functionality of the device.
How to lock the volume control
• Make sure the Multi Mic is turned OFF.
• Press and hold the volume ‘-‘ button and the Power ON but-
ton for 1.5 seconds until the status LED turns green (Multi Mic
powering up) and release the Power button.
• Keep pressing the volume ‘-‘ button until a green-orange-red
LED sequence is shown on the status LED after 10 seconds.
The status LED will now blink twice every 2 seconds to indicate
that the volume control is locked.
How to unlock the volume control
• Make sure the Multi Mic is turned OFF.
• Press and hold the volume ‘+‘ button and the Power ON but-
ton for 1.5 seconds until the status LED turns green (Multi Mic
powering up) and release the Power button.
• Keep pressing the volume ‘+‘ button until a green-orange-red
LED sequence is shown on the status LED after 10 seconds.
The volume can be adjusted independently for Mic and FM mode.
When locked, the volume levels for both Mic mode and FM mode