EverFlo
Q

Questions and Answers

  • K
    Kelly WilliamsAug 18, 2025
    How to fix impeded oxygen flow indication on Respironics EverFlo Medical Equipment?
    • A
      Anthony HodgeAug 18, 2025
      If the Respironics Medical Equipment impeded oxygen flow indication is activated (Yellow LED illuminates continuously, the Red LED is blinking and the Audible Alarm is beeping), first, remove any items blocking airflow to the device. Then, ensure the flow meter knob isn't completely closed; turn it counterclockwise to center the ball on the prescribed LPM flow. Finally, check the oxygen tubing for kinks or blockages and replace it if necessary.
  • M
    Melissa PerryAug 20, 2025
    What to do if Respironics EverFlo is beeping and all LEDs are illuminated?
    • A
      Alison ThomasAug 20, 2025
      If your Respironics Medical Equipment isn't working when turned on, and you hear beeping with all 3 LEDs illuminated, connect to a backup oxygen source and contact your home care provider. This indicates an internal part failure.
  • L
    Latoya LeeAug 23, 2025
    What to do if the Audible Alarm is sounding continuously and none of the LEDs are illuminated on my Respironics EverFlo?
    • E
      Elizabeth FordAug 23, 2025
      If the audible alarm is continuously sounding and none of the LEDs are illuminated, it usually means the device is on but not working. This often happens because the device isn't plugged in or there's a power outage. First, check the power outlet and make sure the device is properly plugged in. If the problem persists, connect to a backup oxygen source and contact your home care provider.
  • A
    Austin ClineAug 26, 2025
    Why is the Red LED continuously illuminated and the Audible Alarm constantly sounding on my Respironics EverFlo?
    • Z
      zleeAug 26, 2025
      If the red LED is continuously lit and the audible alarm is constantly sounding, it indicates the device has detected a system malfunction. You should immediately turn off the device and wait 5 minutes. Then, restart it. If the problem continues, turn the unit off, connect to a backup oxygen source, and call your home care provider.
  • M
    Matthew IbarraAug 28, 2025
    How to fix my Respironics EverFlo if it's not working when turned on, the Audible Alarm is sounding continuously, and all LEDs are off?
    • C
      Crystal HuertaAug 28, 2025
      If your device isn't working when turned on, the audible alarm is sounding continuously, and all LEDs are off, here are a few things to check: * Make sure the power cord is properly inserted into the electrical outlet. * Check your household outlet fuse or circuit to ensure the unit is receiving power. If these steps don't resolve the issue, it may indicate an internal part failure, so connect to a backup oxygen source and contact your home care provider.
  • J
    Jeffrey MeyerAug 30, 2025
    What should I do if all 3 LEDs are continuously illuminated and the Audible Alarm is constantly sounding on my Respironics EverFlo Medical Equipment?
    • K
      Kimberly LopezAug 31, 2025
      If all three LEDs are continuously illuminated and the audible alarm is constantly sounding, the device has detected a system malfunction. Immediately turn off the device, connect to a backup oxygen source, and call your home care provider.
  • D
    Derek DillonSep 2, 2025
    How do I resolve an impeded oxygen flow indication on my Respironics Medical Equipment (Yellow LED is continuously illuminated, the Red LED is blinking, and the Audible Alarm is beeping periodically)?
    • E
      Eric CannonSep 2, 2025
      If you're experiencing an impeded oxygen flow, indicated by the yellow LED being continuously illuminated, the red LED blinking, and the audible alarm beeping periodically, consider the following: * Remove any items that might be blocking airflow into the device. * Ensure the flow meter knob is not completely closed; turn it counterclockwise to center the ball on your prescribed LPM flow. * Check the oxygen tubing for kinks or blockages and replace it if necessary.
  • S
    Sandra PageSep 4, 2025
    Why is my Respironics EverFlo not working when turned on, the Audible Alarm is sounding continuously, and all 3 LEDs are illuminated?
    • J
      John HughesSep 4, 2025
      If your device isn't working when turned on, the audible alarm is sounding continuously, and all 3 LEDs are illuminated, it indicates an internal part failure. In this case, connect to a backup oxygen source and contact your home care provider.
  • D
    Debbie HurstSep 7, 2025
    What does it mean if the Yellow LED illuminates continuously, the Red LED is blinking, and the Audible Alarm is beeping periodically on my Respironics EverFlo?
    • S
      Stanley BatesSep 7, 2025
      If the yellow LED illuminates continuously, the red LED is blinking, and the audible alarm is beeping periodically, it means the device has detected an impeded oxygen flow condition. Consult the troubleshooting guide. Connect to a backup oxygen source and call your home care provider if troubleshooting doesn't resolve the issue.
  • D
    David ScottSep 9, 2025
    What should I do if the Yellow LED is blinking, the Red LED is off, and the Audible Alarm is beeping periodically on my Respironics EverFlo Medical Equipment?
    • C
      Cynthia HobbsSep 10, 2025
      If the yellow LED is blinking, the red LED is off, and the audible alarm is beeping periodically, the device has detected a high oxygen flow condition. Turn the flow rate down to your prescribed level and wait at least 2 minutes. If the condition persists, turn the unit off, connect to a backup oxygen source, and call your home care provider.

Summary

Symbol Key

Abbreviations

Chapter 1: Introduction

Intended Use

Defines the purpose and application of the EverFlo/EverFlo Q Oxygen Concentrator.

About Your EverFlo; EverFlo Q

Provides an overview of the device's function and technology for oxygen concentration.

Parts of Your Concentrator

Identifies the main components and physical parts of the oxygen concentrator device.

Accessory Equipment and Replacement Parts

Lists recommended accessories and replacement parts for the oxygen concentrator.

Warnings and Cautions

Warnings

Details potential hazards to the operator or patient, requiring careful adherence to guidelines.

Cautions

Highlights potential risks of damage to the equipment and necessary precautions.

Chapter 2: Operating Instructions

Chapter 3: Cleaning & Maintenance

Cleaning

Instructions for periodic cleaning of the EverFlo/EverFlo Q device exterior.

Service

Information regarding user-serviceable parts and professional servicing requirements.

How to Contact Respironics

Contact details for customer service and support for the device.

Chapter 4: Alarms and Troubleshooting

Alarm and Indicators

Details the device's audible alarm and three LED indicators for status monitoring.

Troubleshooting Guide

Chapter 5: Specifications

Appendix A: EMC Information

Electromagnetic Emissions

Guidance on electromagnetic emissions and compliance for the device.

Electromagnetic Immunity

Guidance on electromagnetic immunity and compliance for the device.

Recommended Separation Distances

Limited Warranty

Respironics EverFlo Specifications

General IconGeneral
Device TypeOxygen Concentrator
Oxygen Concentration93% ± 3%
Weight31 lbs (14 kg)
Operating HumidityUp to 95% non-condensing
Outlet Pressure5.5 PSI
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Warranty3 years
Flow Rate0.5 - 5 lpm
Power Consumption350 W
Operating Temperature12°C to 32°C (55°F to 90°F)
AltitudeUp to 7, 500 feet
Power Requirements60 Hz
Sound Level45 dB(A)