RESPIRÓNICS"
Chapter
5:
Verification
Procedures
5.3
System
Pressure
Test
Procedure:
Yellow
Pressure
The
system
pressure
test
is
used
to
verify
the
internal
operating
Tubing
pressures
of
the
Millennium
Oxygen
Concentrator.
Step
1
Run
the
unit
with
the
cabinets
on
for
a
minimum
of
one
hour,
Step
2
Remove
the
front
and
rear
cabinets
from
the
unit
(see
Sieve
Sections
8.4.4
and
8.4.5
for
more
detailed
instructions).
Cannister
Fitting
Step
3
Disconnect
the
yellow
pressure
tubing
from
the
fitting
on
the
top
of
the
sieve
cannister
assembly.
Using
the
pressure
gauge,
“T”
fitting,
and
short
yellow
tubing
supplied
with
the
Millennium
tool
kit,
connect
a
pres-
sure
gauge
to
the
long
yellow
pressure
tubing
and
the
,
sieve
canister
assembiy.
Figure
5-2
Connecting
the
Pressure
Gauge
Step
4
Tum
the
unit on
and
set
the
flow
to
5
Ipm.
Allow
the
unit
to
cycle
for
at
least
2
minutes
to
stabilize.
Step
5
Hold
the
pressure
gauge
in
a
vertical
position.
Step
6
Monitor
the
pressure
gauge.
Record
the
peak
pressure
for
four
cycles.
1002403
Millennium
Oxygen
Concentrator
System
Service
Manual
5-5