RIGOL Chapter 2 Calibration Process
2-10 DG1000Z Calibration Guide
9. Use the knob to select the next calibration point (for each calibration point, set a
proper frequency calibration factor on the power meter). Repeat steps 7 and 8
to measure the high-frequency flatness of this calibration point until finishing
the calibration of all the calibration points.
10. Press Save and the high-frequency flatness calibration is finished.