Chapter 1 Test Overview RIGOL
DG1000Z Performance Verification Guide
1-1
Chapter 1 Test Overview
Test Preparations
Before performing the test, make sure that the instrument is within the calibration
period (the recommended calibration period is 1 year) and has been warmed up for
at least 30 minutes under the specified operation temperature (18 to 28).
Recommended Test Devices
It is recommended that you use the test devices listed in the table below or other test
devices whose performance specifications satisfy the Performance Requirement
listed in the table below to test the performance specifications of the DG1000Z
series.
Table 1-1 Recommended test devices
Device Performance Requirement
Recommended
Instrument
Frequency Counter
>10MHz
Accuracy: 0.1ppm
Agilent 53131A
Digital Multimeter 6
1
/
2
digits
RIGOL DM3068
Power Meter
-30dBm to +20dBm
Accuracy: ±0.02dB
Resolution: 0.01dB
Agilent E4418B
Spectrum Analyzer
Minimum resolution bandwidth is
100Hz
RIGOL DSA815
Oscilloscope
Bandwidth: 500MHz
Rise/Fall time measurement
function
Overshoot measurement
function
RIGOL DS4000 series
Connecting Cable BNC (m)-BNC (m)
--
Connecting Cable BNC (m)-Dual banana plug (m)
--