RIGOL Chapter 2 Performance Verification Test
DG1000Z Performance Verification Guide
2-20
Overshoot Test
Specification:
Signal Characteristic
Square
Overshoot
Typical (100kHz, 1Vpp)
≤5%
Test Procedures:
1. Make sure that the environment temperature is between 18 and 28 and
DG1000Z has been warmed up for at least 30 minutes. Connect the channel
output terminal (take CH1 as an example; the test method is also applicable to
CH2) of DG1000Z with the signal input terminal of the oscilloscope using a
dual-BNC connecting cable as shown in Figure 2-5.
2. Turn on DG1000Z. Press Utility Set To Default OK to restore DG1000Z
to factory setting.
3. Set DG1000Z:
a) Set the output impedance of CH1 to 50Ω (press Utility Channel Set
Output Set Imped and select Load).
b) Set the output waveform of CH1 to a square waveform with 100kHz
frequency, 1Vpp amplitude and 0V
DC
offset.
c) Press Output1 to turn on the output of CH1.
4. Turn on and set the oscilloscope:
a) Set the input impedance to 50Ω.
b) Set the vertical scale to 200mV/div.
c) Set the horizontal time base to 50ns.
d) Adjust the trigger level to a proper value.
e) Turn on the overshoot measurement function.
5. Record the overshoot measurement value and judge whether it is within the
specification range.