RIGOL Chapter 2 Performance Verification Test
DG1000Z Performance Verification Guide
2-18
Rise/Fall Time Test
Specification:
Signal Characteristic
Square
Rise/Fall Time
Typical (1Vpp)
<10ns
Test Procedures:
1. Make sure that the environment temperature is between 18℃ and 28℃ and
DG1000Z has been warmed up for at least 30 minutes. Connect the channel
output terminal (take CH1 as an example; the test method is also applicable to
CH2) of DG1000Z with the signal input terminal of the oscilloscope using a
dual-BNC connecting cable as shown in Figure 2-5.
DG1000Z Series Oscilloscope
Figure 2-5 Connect DG1000Z and the Oscilloscope
2. Turn on DG1000Z. Press Utility Set To Default OK to restore DG1000Z
to the factory setting.
3. Set DG1000Z:
a) Set the output impedance of CH1 to 50Ω (press Utility Channel Set
Output Set Imped and select “Load”).
b) Set the output waveform of CH1 to a square waveform with 1MHz frequency,
1Vpp amplitude and 0V
DC
offset.
c) Press Output1 to turn on the output of CH1.
◎