RIGOL Chapter 2 Performance Verification Test
DG1000Z Performance Verification Guide
2-18
Rise/Fall Time Test
Specification:
Signal Characteristic
Square
Rise/Fall Time
Typical (1Vpp)
<10ns
Test Procedures:
1. Make sure that the environment temperature is between 18 and 28 and
DG1000Z has been warmed up for at least 30 minutes. Connect the channel
output terminal (take CH1 as an example; the test method is also applicable to
CH2) of DG1000Z with the signal input terminal of the oscilloscope using a
dual-BNC connecting cable as shown in Figure 2-5.
DG1000Z Series Oscilloscope
Figure 2-5 Connect DG1000Z and the Oscilloscope
2. Turn on DG1000Z. Press Utility Set To Default OK to restore DG1000Z
to the factory setting.
3. Set DG1000Z:
a) Set the output impedance of CH1 to 50Ω (press Utility Channel Set
Output Set Imped and select Load).
b) Set the output waveform of CH1 to a square waveform with 1MHz frequency,
1Vpp amplitude and 0V
DC
offset.
c) Press Output1 to turn on the output of CH1.