RIGOL Chapter 2 Performance Verification Test
DG1000Z Performance Verification Guide
2-12
Harmonic Distortion Test
Specification:
Sine Wave Spectrum Purity
Harmonic Distortion
Typical (0dBm)
DC to 10MHz (include 10MHz): <-65dBc
10MHz to 30MHz (include 30MHz): <-55dBc
30MHz to 60MHz (include 60MHz): <-50dBc
Test Procedures:
1. Make sure that the environment temperature is between 18 and 28 and
DG1000Z has been warmed up for at least 30 minutes. Connect the channel
output terminal (take CH1 as an example; the test method is also applicable to
CH2) of DG1000Z with the signal input terminal of the spectrum analyzer using a
dual-BNC connecting cable and N-BNC adaptor as shown in Figure 2-4.
DG1000Z Series Spectrum Analyzer
Figure 2-4 Connect DG1000Z and the Spectrum Analyzer
2. Turn on DG1000Z. Press Utility Set To Default OK to restore DG1000Z
to the factory setting.
3. Set DG1000Z:
a) Set the output impedance of CH1 to 50Ω (press Utility Channel Set
Output Set Imped and select Load).
b) Set the output waveform of CH1 to a sine waveform with 10MHz frequency,
0dBm amplitude and 0V
DC
offset.
c) Press Output1 to turn on the output of CH1.