EasyManua.ls Logo

Rigol DG1032Z - Page 30

Rigol DG1032Z
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Performance Verification Test
DG1000Z Performance Verification Guide
2-14
measurement value
For example, when the output waveform frequency of the channel is 10MHz, if the base
waveform measurement value is 0.8dBm and the 2
nd
order harmonic measurement value
is -66.2dBm, the 2
nd
order harmonic distortion = (-66.2) -0.8=-67dBc<-65dBc and the
test result fulfills the specification requirement.
Test Record Form:
Frequency
Setting
Value
Setting
Measuremen
t Value
Calculation
Result
[1]
Specification Pass/Fail
10MHz
Waveform:
Sine
Amplitude:
0dBm
Offset: 0V
DC
Base
waveform:
<-65dBc
2
nd
order
harmonic:
25MHz
Base
waveform:
<-55dBc
2
nd
order
harmonic:
30MHz
Base
waveform:
<-55dBc
2
nd
order
harmonic:
60MHz
Base
waveform:
<-50dBc
2
nd
order
harmonic:
Note
[1]
: Calculation result = 2th order harmonic measurement value - base waveform measurement value.