Parameter
Name Type Range Default
-- Keyword NORMal|INVerted NORMal
Explanation
When the output polarity is set to INVerted, the waveform inverts relatively to
the offset voltage.
After the waveform is inverted, none of the offset voltages would change, the
waveform displayed in the user interface is not inverted and the related sync
signal is not inverted.
Return Format
The query returns NORMAL or INVERTED.
Example
The command below sets the output polarity of the [Output1] connector to
INVerted.
:OUTPut1:POLarity INVerted
The query returns INVERTED.
:OUTPut1:POLarity?