Parameter
Name Type Range Default
<percent>
Consecutive Real
Number
0% to 50% 20%
Explanation
The duty cycle deviation <percent> fulfills the following requirements:
The duty cycle deviation can not exceed the duty cycle of the current pulse.
The duty cycle deviation is limited by the minimum duty cycle and the current edge
time.
Return Format
The query returns the duty cycle deviation in scientific notation.
Example
The command below sets the duty cycle deviation to 45%.
:MOD:PWM:DCYCle 45%
The query below returns 4.500000E+01.
:MOD:PWM:DCYCle?