Parameter
Name Type Range Default
<seconds>
Consecutive Real
Number
Refer to the Explanation
below
1.9531 μs
Explanation
The range available is limited by the pulse width currently specified. The relation
fulfills the inequality: leading/falling edge time ≤ 0.625 × pulse width.
DG4000 will automatically adjust the edge time to match the specified pulse
width if the value currently set exceeds the limit value.
Return Format
The query returns the time in scientific notation and the default unit is "s".
Example
The command below sets the leading edge time to 10 μs.
:PULSe:TRANsition 0.00001
The query below returns 1.000000E-05.
:PULSe:TRANsition?