EasyManua.ls Logo

Rigol DG800 Series - Page 66

Rigol DG800 Series
209 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Command System
2-52 DG800 Programming Guide
digits and the default units are Hz, Vpp, V
dc
and ° respectively; the absent item is replaced by DEF) of the
waveform of the specified channel respectively. For example,
"SQU,1.000000E+03,2.000000E+00,3.000000E+00,4.000000E+00" denotes that the current waveform is
Square, the frequency is 1 kHz, the amplitude is 2 Vpp, the offset is 3 V
dc
, and the start phase is 4°.
Example
:SOUR1:APPL? /*Queries the waveform type as well as the frequency, amplitude, offset, and phase of CH1;
the query returns SQU,1.000000E+03,2.000000E+00,3.000000E+00,4.000000E+00".*/
[:SOURce[<n>]]:APPLy:DC
Syntax
[:SOURce[<n>]]:APPLy:DC
[{<frequency>|DEF}[,{<amplitude>|DEF}[,{<offset>|DEFault|MINimum|MAXimum}]]]
Description
Sets the waveform of the specified channel to DC with the specified offset.
Parameter
Name
Type
Range
Default
[<n>]
Discrete
1|2
1
<offset>
Real
Refer to "Remarks".
0 V
dc
Remarks
When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default.
<frequency> and <amplitude> are not applicable to the DC function, but a placeholder must be
specified for them.
The range of <offset> is limited by the "Impedance" (
:OUTPut[<n>]:IMPedance
or
:OUTPut[<n>]:LOAD) setting.
Example
:SOUR1:APPL:DC 1,1,2 /*Sets the waveform of CH1 to DC, with 2 V
dc
offset.*/
[:SOURce[<n>]]:APPLy:DUALTone
Syntax
[:SOURce[<n>]]:APPLy:DUALTone
[<freq>|DEFault|MINimum|MAXimum[,<amp>|DEFault|MINimum|MAXimum[,<offset>|DEFault|MINimu
m|MAXimum]]]
Description
Sets the waveform of the specified channel to Dual-tone waveform with the specified frequency, amplitude,
offset, and phase.
Parameter
Name
Type
Range
Default
[<n>]
Discrete
1|2
1
<freq>
Real
1 μHz to 20 MHz
1 kHz
<amp>
Real
Refer to "Remarks".
5 Vpp
<offset>
Real
Refer to "Remarks".
0 V
dc
Remarks
When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default.
The range of <amp> is limited by the "Impedance" (
:OUTPut[<n>]:IMPedance
or
:OUTPut[<n>]:LOAD) and "Frequency/Period" settings. The range of <offset> is limited by the