Chapter 2 Command System  RIGOL 
DG800 Programming Guide  2-71 
Parameter 
<frequency>  Real 
Frequency range of the current 
waveform of the specified channel 
900 Hz 
 
Remarks   
  You can set the sweep boundaries via the center frequency and frequency span 
(
[:SOURce[<n>]]:FREQuency:SPAN). Different sweep waveforms correspond to different center 
frequency and frequency span ranges. Besides, the center frequency and frequency span are mutually 
affected. Define the minimum frequency of the currently selected waveform
, the maximum 
frequency as 
 and 
. The range of the center frequency (defined as 
) 
is from 
 to 
. The range of the frequency span (defined as 
) is affected by the center 
frequency. When the center frequency is smaller than 
, the range of the frequency span is 
; when the center frequency is greater than 
, the range of the frequency span is 
. 
  When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default. 
 
The start frequency, stop frequency, center frequency and frequency span fulfills the following 
equations.
 
|)(|
stopstartcenter
FFF +=
;
 
Wherein, 
  ——center frequency; 
  ——frequency span; 
  ——start frequency; 
  ——stop frequency. 
  The generator will restart to sweep from the specified "start frequency" after the "frequency span" is 
modified. In large-scale sweep, the amplitude characteristic of the output signal might change.
 
 
Return Format 
The query returns the frequency span in scientific notation, with 7 effective digits. For example, 
8.000000E+02 (the frequency span is 800 Hz). 
 
Example 
:SOUR1:FREQ:SPAN 800    /*Sets the frequency span of the sweep function of CH1 to 800 Hz.*/ 
:SOUR1:FREQ:SPAN?      /*Queries the frequency span of the sweep function of CH1 and the query 
returns 8.000000E+02.*/ 
 
 
[:SOURce[<n>]]:FREQuency:STARt 
 
Syntax 
[:SOURce[<n>]]:FREQuency:STARt {<frequency>|MINimum|MAXimum} 
[:SOURce[<n>]]:FREQuency:STARt? [MINimum|MAXimum] 
 
Description 
Sets the start frequency of the sweep function of the specified channel. 
Queries the start frequency of the sweep function of the specified channel. 
 
Parameter 
<frequency>  Real 
Frequency range of the current 
waveform of the specified channel 
100 Hz