Chapter 2 Performance Verification Test RIGOL
DM3058/DM3058E Performance Verification Guide
Note
[1]
: The reading rate is set to "Slow".
Note
[2]
: For continuous currents greater than DC 7A or AC RMS 7A, close the circuit for 30 s and
then open the circuit for 30 s.
Note
[3]
: Specifications are for the 4-wire resistance function or the 2-wire resistance function using
“REL” operation. If the 2-wire resistance function is used without "REL" operation, add 0.2 Ω
additional errors.
Note
[4]
: Use the 2-wire resistance function.
Note
[5]
: Q marks the optional quick test points.
2. The AC voltage gain verification procedures
(1) Make sure you have carefully read the "Test Conditions".
(2) Connect the input terminals of the multimeter with the calibrator.
(3) Test the ranges specified in Table 2-3 one by one. Set the reading rate to
"Slow" and turn off all the math operations.
(4) Input the signals specified in Table 2-3 using the calibrator and compare the
test results with the test limits in the table (make sure the output of the
calibrator is built adequately).
Table 2-3 AC Voltage Gain Verification Table
Range
[1]
Test
Signal
Input
Frequency
Quick
Test
1-year Test Limit
Upper Limit Lower Limit
200.000 mV
200.000 mV
200 mV 50 kHz 202.100 mV 197.900 mV
2.00000 V
2 V 20 kHz Q 2.00500 V 1.99500 V
2 V 50 kHz Q 2.02100 V 1.97900 V
2 V 100 kHz Q 2.06100 V 1.93900 V
20.0000 V
20 V 100 kHz 20.6100 V 19.3900 V