Chapter 1 Test Overview RIGOL
DP800 Performance Verification Manual
1-3
4. Make sure that the environment temperature is between 20 and 30 and the
relative humidity is less than 80%.
5. The test connecting wires used should be as short as possible. For the channels
that support the Sense function, please short-circuit the (S+) and (+) terminals,
(S-) and (-) terminals (as shown in Figure 2-2 or Figure 3-2) of the channels
respectively and turn on the Sense function of the corresponding channel before
performing the test.
The tests introduced in this manual should be performed by professionals. During
the tests, there might be dangerous voltage at the output terminals of the power
supply as well as certain test devices.