EasyManuals Logo

Rigol DP832 Performance Verification Manual

Rigol DP832
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
Appendix RIGOL
DP800 Performance Verification Manual
17
DP811A Performance Verification Test Record Form
RIGOL DP811A Programmable Linear DC Power Supply
Performance Verification Test Record Form
CV Load Regulation Rate
Range Range1 (20V/10A) Range2 (40V/5A)
Specification <4mV <6mV
DP811A
Setting
Voltage 20V 40V
Current 10.5A 5.3A
Measurement
Result
U
0
U
1
|U
1
-U
0
|
CV Linear Regulation Rate
Range Range1 (20V/10A) Range2 (40V/5A)
Specification <1.25mA
<750μA
DP811A
Setting
Voltage 21V 42V
Current 10A 5A
Measurement
Result
U
0
U
1
U
2
|U
1
-U
0
|/R
M
[1]
|U
2
-U
0
|/R
M
[1]
CV Ripple and Noise
Range Range1 (20V/10A) Range2 (40V/5A)
Specification <350μVrms/3mVpp
Resistive Load R
L
2Ω 8Ω
DP811A
Setting
Voltage 20V 40V
Current 10.5A 5.3A
Measurement
Result
Peak-peak value
(Vpp)
RMS value
(Vrms)

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232
Display3.5 inch TFT LCD