Chapter 2 Constant Voltage Tests RIGOL
DP800 Performance Verification Manual
2-3
Note:
During the CV tests, make connections according to Figure 2-1 for the normal
channels (namely the channels that do not support the Sense function) of the DP800
series power supply. For the channels that support the Sense function (CH2 of
DP821A and DP811A), make connections according to Figure 2-2 (note to
short-circuit the (S+) and (+) terminals, (S-) and (-) terminals of the channel
respectively) and turn on the Sense function of the channel during the test process to
measure the specifications under Sense mode.