EasyManua.ls Logo

Rigol DP832 - Constant Current Output

Rigol DP832
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Constant Voltage Tests
DP800 Performance Verification Manual
2-24
Table 2-23 DP832A CV Programming and Readback Accuracy Test Record Form
Channel CH1 (30V/3A) CH2 (30V/3A) CH3 (5V/3A)
Specification
Programming <35mV <35mV <10mV
Readback <25mV <25mV <10mV
The output voltage is 0
DP832A
Setting
Voltage 0V 0V 0V
Current 3.2A 3.2A 3.2A
Measurement
Result
U
1
U
2
Programming:
|U
1
-0|
Readback:
|U
2
-U
1
|
The output voltage is U
R
[1]
DP832A
Setting
Voltage 30V 30V 5V
Current 3.2A 3.2A 3.2A
Measurement
Result
U
3
U
4
Programming:
|U
3
-U
R
[1]
|
Readback:
|U
4
-U
3
|
Note
[1]
: The rated voltage of the specified channel.

Other manuals for Rigol DP832