RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-18
17) Repeat step 6).
18) Output a sine waveform with 20 MHz frequency and 1.2 Vpp amplitude from
Fluke 9500B.
19) Repeat step 8), 9) and 10).
20) Output a sine waveform with 50 MHz frequency and 1.2 Vpp amplitude from
Fluke 9500B.
21) Repeat step 12), 13) and 14).
22) Keep other settings of the oscilloscope unchanged and set the vertical scale
to 500 mV/div.
23) Output a sine waveform with 1 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
24) Repeat step 6).
25) Output a sine waveform with 20 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
26) Repeat step 8), 9) and 10).
27) Output a sine waveform with 50 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
28) Repeat step 12), 13) and 14).
29) Turn CH1 off. Test CH2 according to the above test steps and record the test
results.

Table of Contents

Other manuals for Rigol DS1052E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1052E and is the answer not in the manual?

Rigol DS1052E Specifications

General IconGeneral
Bandwidth50 MHz
Sample Rate1 GSa/s
Channels2
Memory Depth1 Mpts
Vertical Sensitivity2 mV/div to 10 V/div
Input Impedance1 MΩ ± 2% || 15 pF ± 3 pF
Vertical Resolution8 bits
Trigger ModesEdge, Pulse, Video, Slope, Alternate
Power Supply100-240 V AC, 45-440 Hz
Display5.7 inch TFT LCD
InterfacesUSB Host, USB Device, RS-232
Dimensions303 mm x 154 mm x 133 mm

Related product manuals