Command Systems        RIGOL 
 
Programming Guide for DS1000B Series 
Returned Format:   
The query returns DC, AC or LF. 
 
Example: 
:TRIG:COUP DC     Set the coupling mode as DC. 
:TRIG:COUP?    Return DC. 
 
 
7.  :TRIGger:HFREject 
Command Format:  
:TRIGger:HFREject {{1|ON}|{0|OFF}} 
:TRIGger:HFREject? 
 
Function: 
The command sets high frequency reject function on or off. 
 
Returned Format:   
The query returns 1 or 0, respectively indicates ON or OFF. 
 
Example: 
:TRIG:HFRE ON      Set HFR on. 
:TRIG:HFRE?     Return 1. 
 
 
8.  :TRIGger:HOLDoff 
Command Format:  
:TRIGger:HOLDoff <count> 
:TRIGger:HOLDoff? 
 
Function: 
The command sets the holdoff time of a trigger. Holdoff time indicates the 
waiting time before oscilloscope starts a new trigger. During Holdoff, the 
oscilloscope will not trigger until Holdoff ends. 
The <count> range is 100ns~1.5s. 
 
Returned Format:   
The query returns the value of holdoff time, and the unit is s.