Command Systems        RIGOL 
 
Programming Guide for DS1000B Series 
:TRIG:ALT:TSCAL? SOURB    Return 1.000e-003. 
 
 
5.  :TRIGger:ALTernation:TimeOFFSet 
Command Format:  
:TRIGger:ALTernation:TimeOFFSet <value>[,<source>] 
:TRIGger:ALTernation:TimeOFFSet? [<source>] 
 
Function: 
The command sets the timebase offset. 
In NORMAL mode, <value>: 1s ~ memory capacitance; 
In STOP mode, <value: -500s ~ +500s. 
     
Returned Format:   
The query returns the value of timebase offset, and the unit is s.  
     
Example: 
:TRIG:ALT:TOFFS 0.0002,SOURB   Set the timebase offset as 200μs. 
:TRIG:ALT:TOFFS? SOURB     Return 2.000e-004. 
 
 
6.  :TRIGger:ALTernation:LEVel 
Command Format:  
:TRIGger:ALTernation:LEVel <value>[,<source>] 
:TRIGger:ALTernation:LEVel? [<source>] 
 
Function: 
The command sets the trigger level of current channel. The <value> range  
<value> range: (- 6*Scale-Offset
①) ~(+6*Scale+Offset①).Scale is the current 
vertical scale, and the unit is V/div. the <source> may be SOURceA or SOURceB, 
and the source A and B is different according to the current alternation channel. 
NOTE①: Trigger Level range is up to+/-6 Scale, when channel has offset, it needs to detract 
offset ,such as 1V tap position, 1V offset, the trigger range is -7V~5V.     
 
Returned Format:   
The query returns the value of trigger voltage level, and the unit is V.