Command Systems        RIGOL 
 
Programming Guide for DS1000B Series 
 
5.  :WAVeform:SOURce 
Command Format:  
:WAVeform:SOURce <source> 
:WAVeform:SOURce? 
 
Function: 
The command sets the source of waveform data which is going to check. The 
<source> may be CHANnel1, CHANnel2, CHANnel3, CHANnel4 or MATH. 
Returned Format:   
The query returns Channel1 or Channel2, Channel3, Channel4, MATH. 
 
Example: 
:WAV:SOUR CHAN2   Set the data source as channel 2. 
:WAV:SOUR?       Return Channel2. 
 
 
6.  :WAVeform:PREamble? 
Command Format: 
  :WAVeform:PREamble? 
 
  Function: 
This command queries the current waveform settings. 
 
Returned Format: 
The query returns 10 data which are separated by comma “,”, they are: 
Format,Type,Points,Count,Xinc,Xor,Xref,Yinc,Yor,Yref 
 
Parameter Value: 
Format: BYTE – 0; WORD – 1; ASCII – 2. 
Type: NORMAL – 0; PEAK_DETECT – 1; AVERAGE – 2. 
Points: specified by 
:WAVeform:POINts command. 
Count: the “average acquisition time” (in average mode) or “1” (other mode); 
Xinc: 1/SaRate (RAW) or TimeScale/50 (NORMAL); 
Xor: relative time of the trigger points; 
Xref: X reference; 
Yinc: Y unit voltage;