RIGOL Appendix Test Record Form
DS6000 Performance Verification Guide
4
(Continue) 1 MΩ Input Impedance Test Record Form
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2 Calculation Result
[1]
CH4
2 mV/div
2%
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
Note
[1]
: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale × 100%; wherein, Vout1 and Vout2
are 3 × the current vertical scale and -3 × the current vertical scale respectively.
Bandwidth Test
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vrms1 Vrms2 Amplitude Loss
[1]
CH1
200 mV/div
-3 dB to 1 dB
500 mV/div
CH2
200 mV/div
500 mV/div
CH3
200 mV/div
500 mV/div
CH4
200 mV/div
500 mV/div
Note
[1]
: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).